If the devices shows as "PI USB to Serial" then use the below file named If the devices shows as "Gadget Serial v2.4" then use the below file named usbser.sys If it does not automatically do this, download one of the two driver files attached to this article (links at bottom of article) and save to this path, C:\Windows\System32\drivers and then have Windows search for it again, or manually update the driver from Windows Device Manager. With the above settings, when you plug the 6270A into a computer with a USB cable the computer should recognize that a new device is connected and try to load/find the driver for it. Ensure that Emulation Mode is set to SCPI If "RS232 Emulation" is set to Off then the controller can be seen like a USB memory stick on the computer (transfer saved image files from controller to computer).Ĭheck the baud rate, and other settings on the Remote Port screen by clicking on. If "Remote IF" is set to Terminal then every command to the 6270A would be echoed back to the computer. To use the USB connection for RS232 communications follow this menu path on the 6270A touch screen: Setup -> Instrument Setup -> Remote Port -> USB SetupĮnsure that " Remote IF" is set to Computerand " RS232 Emulation" is set to ON. This works the same on a 2271A, 8270A or 8370A pressure controller also.ġ. The 6270A Pressure Controller's USB port on the back panel can be plugged into a computer and then the 6270A can be communicated with with RS232 communications (or files can be transferred to/from the 6270A like it's an external drive).